Apply the SpO
-Sensor
2
cable protection
1
SpO
– Sensor
2
2
pad
3
Fig. 24
Apply the SpO
Press the two pads (3) together and push the opened sensor on any finger, so that
the cable / socket side of the sensor is located on the finger nail side.
Reusable sensors may remain at one measuring point for 4 h maximally, provided
that state of the skin, the correct and secure positioning of the sensor at the measur-
ing point are controlled regularly. As the tolerance towards sensors at the measuring
point depends upon the individual condition of the skin, it may be necessary to
change the points of measure with some patients.
Connect the SpO
Connect the SpO
-Sensor plug (1) into the socket of the device, the arrow (3) at
2
the plug and the arrow (4) at the socket must match together. Make sure that the
plug has been connected completely before you use it.
To disconnect the connector, lift the sleeve with arrow first up and then pull the
connector out from the socket.