The AUTO-SYNC is only available in the Manual Mode and is only carried out on
lead II.
In case of synchronous operation the release button has to be pressed down until the
moment for the cardioversion is reached. During this time an intermittent signal will
be on. No cardioversion will be released, if during discharging the release buttons are
let go off. The energy is discharged internally if no synchronisation takes place within
3 seconds while keeping the release button pressed down.
If the activation button is only pressed briefly and if the defibrillation is not triggered
by synchronised markers, the energy remains in the capacitor.
The energies required for synchronized cardioversion are in most cases lower than for
defibrillation (unsynchronized cardioversion) as it is not necessary to depolarize all
heart muscles.
The energy depends roughly on the patient's body height and weight. The indica-
tions, however, determine the energy, i. e. the following empirical values are applica-
ble:
Ventricular tachycardia with un-
stable pulse:
Supraventricular tachycardia:
Atrial flutter:
Atrial fibrillation:
The above mentioned values are only recommendations.

Keeping The Defibrillator Ready For Use

Clean the device after resuscitation, replace the SavePads and check and / or change
the power supply unit if required, so that PRIMEDIC™ DefiMonitor can be used as
fast as possible on the next use. Charge the AkuPak for sufficient power on the next
use.
Contact the nearest service station in case of any faults or irregularities as soon as
possible.
50 joule, further cardioversions to be increased
by approx. 50 joule each (100 j, 200 j, ...)