Conditions For The Start Of The Analysis; Automatic Start Of Analysis; Adapting The Volume Of The Metronome - Primedic HeartSave Operating Instructions Manual

Aed trainer
Table of Contents

Advertisement

Using the device
Note
The Voice Announcement < Chest compressions > during the energy charging
is omitted in Pediatric Mode.

4.2.3 Conditions for the Start of the Analysis

By using training pads on electrically non-conductive resuscitation dolls the PRIMEDIC ™
Heart Save AED Trainer cannot automatically start the analysis, when the user electrodes are
adhered to the dolls. Therefore, the instructor must simulate the electrode contact during
training.
 Press the  button on the remote control (impedance on/off) to simulate the first
electrode contact.
The device then automatically starts the analysis, provided the training cord is plugged in.
 Pushing the button  again switches patient impedance off again. This simulates the
electrodes falling off.

4.2.4 Automatic start of analysis

You can set the HeartSave AED Trainer in such a way that the analysis starts automatically. In
this case the patient impedance must not be added using the remote control.
Proceed as follows to activate this function:
 Push the button 7 (3) on the remote control. If the automatic patient impedance is
activated, the HeartSave AED Trainer emits a rising tone sequence.
Proceed as follows to de-activate this function:
 Push the button 7 (3) on the remote control. If the automatic patient impedance is de-
activated, the HeartSave AED Trainer emits a falling tone sequence.
Note
The setting is saved until the energy module is removed.

4.2.5 Adapting the volume of the metronome

You can adapt the volume of the metronome by pushing the button (1) or (8) during the
metronome phase. The volume of the voice messages is not affected by this.
Note
The setting is saved until the energy module is removed.
18 / 24
23560 HeartSave AED Trainer / GB / B02
METRAX GmbH