6.2.1
6.2.2
6.2.3
6.3
21324 / 09.05 / 02 / GB
DefiMonitor XD
Changing the Configuration – Example
On entering the Setup Menu, the first item in the menu is selected.
For example, if you want to update the time, move the selection downwards by
pressing the
button once or more until the item Time is selected.
Confirm the selected menu item Time by pressing the
to hours.
Change the hour by pressing the
Confirm the correct value with the
and then to seconds.
Change these values in the manner described above. After the final confirmation with
the
button, the selection once again jumps to the menu item Time.
Now carry out any other changes (if required) in the same manner.
To exit the Setup Menu, move the selection with the
item END and confirm this with the
PRIMEDIC™ DefiMonitor is then ready for operation again.

Choosing The Operation Mode

The DefiMonitor XD100 / XD110 / XD300 / XD330 offers two different operating
modes:
– The AUTO-Mode and
– The MANUAL Mode

Auto Mode

After switching on and successfully passed self test the device always operates in
Auto Mode.
This operation mode carries out an automatic ECG-Rhythm-Analysis.
If the device clearly detects VF or VT, it recommends a defibrillation and prepares for
it internally and automatically and recommends to deliver a defibrillation to the pa-
tient. If the device does not detect a rhythm requiring defibrillation, it recommends
cardiopulmonary resuscitation.

Manual Mode

By pressing the button MAN/AUTO again the DefiMonitor will be switched to Man-
ual mode. The manual mode is not activated.
Switch to operation mode AUTO
– By switching the device off and on again or
– By pressing the button Man/AUTO.

Alarms

The alarms (heart rate and SpO
the monitor.
The setting range for the lower alarm limit is in the range of 30 -180, with the upper
alarm limit of 60-300. This is displayed in the monitor. The alarms can be activated in
or
button.
button. The selection then jumps to minutes
button.
) are inactive after switching on which is indicated on