Connecting The Powercord

Insert the powercord in the device socket and then insert the powercord into a socket
near the patient. The powercord is around three meters long.
Make sure that the either the cover or the AkuPak are inserted into the power mod-
ule compartment. This covering is mandatory to ensure peferct functioning of the de-
vice during mains power supply use.

Inserting the printer paper

In order to install the printer paper (paper roll) please push the release button above
the paper compartment. The printer cover will open out to the front. Prepare the pa-
per roll (remove tape, unroll paper approx. 5 cm). Insert paper roll into printer com-
partment with the squared side upwards. Close the cover.
The cover is constructed in a way that it will remove automatically without damage
whenever force towards it will be exerted from outside. Place the cover back onto the
initial position until the fixation will be audible.

Device Self Test

Self Test After Switch On Of DefiMonitor

The self test will either be activated when switching on the DefiMonitor or inserting
the energy module into the device and passes immediately the device self test to
check all important functions and signal settings.
If the energy module was changed and the DefiMonitor has located a failure the
automatic self test (FULL) will take place. Follow the instructions of the DefiMoni-
tor.

Automatic, Periodical self test

The device is carrying out automatic self tests in order to ensure full functionality. To
do this, the DefiMonitor must either be connected to the power supplied or charged
battery pack.
Cycle
SHORT
Daily
MEDIUM
First day of the month
LONG
On July 1st and January
1st of each year
DefiMonitor XD
Test
Software, keyboard, battery, ECG
calibration, clock, internal voltage
supply and HV part at 0 V
Software, keyboard, battery, ECG
calibration, clock, internal voltage
supply and HV part at 300 V
Software, keyboard, battery, ECG
calibration, clock, internal voltage
supply and HV part at 1600 V
21324 / 09.05 / 02 / GB