The anterior posterior position has the advantage that the stimulation intensity will be
lower and therefore the cheat muscles will be less stimulated. This has a positive in-
fluence on the physical comfort of the patient.
If the anterior-posterior position is contraindicated (e.g. if the onset of ventricular fib-
rillation is likely), alternatively the anterior - anterior position might be used.
The position corresponds to the paddle position during defibrillation.
Anterior
Blue
( - )
Anterior
Whi-
te
( + )
To prevent the danger of electrocution it is important to pay attention to the follow-
ing order when attaching the adhesive electrodes. First connect the patient cable with
the snap-in connector with the attached electrodes at the patient. Then connect elec-
trode cable with the DefiMonitor.
If a defibrillation is indicated during pacing, defibrillation might be done by with the
Pacer electrodes. Note the electrode position when doing a defibrillation.
Adults may only be paced with respective electrodes for adults, children only with the
respective paediatric electrodes.
Use only original PRIMEDIC

Opening SavePads

Open the packing of the electrodes by tearing open the protective covering.

Removing the Chest Hair

If the patient has hair on the electrodes' positions, then it must be shaved off.
Shave off the hair at the electrodes' positions with the given shaver.

Connecting The Electrodes

Remove the electrodes from the SavePads pack.
Insert the connector of the electrode cable in the socket of theswitched
on DefiMonitor.
Pay attention that the lock engages.
In order to remove the defibrillation electrode plug you must push
firmly on the upper part of the catch and simultaneously pull out the
plug.
Position
respective Apex position, left frontal chest area, over apex
on the auxiliary line
respective Sternum position, right frontal cheat area be-
yond collarbone
Pacer-/Defibrillation electrodes.