Event; Operating The Printer; Log Of The Ecg Signal - Primedic DefiMonitor XD Series Operating Manual

Table of Contents

Advertisement

Operating Manual
6.4
6.5
6.5.1
32
the Setup menue. After acknowledging ALARM press the alarm button that the
acoustical alarm is suppressed for about 1 minute.
Display in the monitor (e.g. ALARM 30 / 90, SpO
frequency is exceeded by 90 beats/minute or if it is less than 30 beats / minute, a
rhythmical alarm sounds. The alarm bell symbol flashes If the alarm sounds you can
acknowledge this (shut off) by briefly pressing the alarm key. The alarm key is shut
off for approx. 1 minute but the alarm limits remain. When the mute setting is acti-
vated, the alarm bell symbol appears crossed-out.
The heart alarm is automatically set to OFF after defibrillation.

Event

A mark is placed by pressing the event key in the EKG which saves the EKG 5 seconds
before and 5 seconds after the event. This event can then be printed out later with
the event memory and/or viewed in the saved patient vile on the PC using the EKG
viewer.

Operating The Printer

The printer can only be used in Manual mode.

Log of the ECG Signal

The PRIMEDIC™ DefiMonitor XD has a printer with high resolution. An EKG pres-
sure with 3 to 6 channels simultaneously is possible. 25 and 50 mm/s print speed is
available.
The online print starts to log the EKS curve during monitoring by pressing the Printer
On/Off key.
Pressing the printer ON/OFF key stops the log printout. The EKG printout occurs with
the parameters which are selected in the setup menu. The following settings are avai-
lable:
1-channel print
3-channel print
6-channel pressure
25 mm/s – print
speed
50 mm/s – print
speed
2
prints the EKG channel shown in the monitor. If the
SpO
measurement is active, the Spo
2
shown.
prints the channels I,II, III or aVR, aVL, aVF simultane-
ously, depending on which derivation is shown in the
monitor.
prints the maximum possible derivations I, II. III, aVR,
aVL, aVF in real time, depending on the connected elec-
trodes, in 3 blocks.
The printout is at 25 mm/s.
The printout is at 50 mm/s.
DefiMonitor XD
90 / 100) means that if the heart
pulse curve is also
2
21324 / 09.05 / 02 / GB