Operation Of The Pacer; Switching On And Off - Primedic DefiMonitor XD Series Operating Manual

Table of Contents

Advertisement

Operating Manual
11
11.1
44

Operation Of The Pacer

Switching On and Off

While using the PRIMEDIC
sidered:
-
Switch PRIMEDIC
DefiMonitor XD on.
-
Apply adhesive electrodes to patient.
-
Connect adhesive electrodes with snap-in connector with the SavePads Plus ca-
ble. This is done by first opening the snap-in connector with the lid. The metal
flap connector of the electrode is inserted into the slit of the snap-in connector.
Close the lid to fixate the connection. Pay attention to the color marking for the
polarity of the respective connector at the electrode position.
-
Connect SavePads Plus cable to the socket at the DefiMonitor.
-
Attach ECG electrodes to the patient and connect ECG patient cable with socket
DefiMonitor.
of PRIMEDIC
-
Switch on PRIMEDIC
The PRIMEDIC
Pacer XD is switched on by a short press of the Pacer On- / Off-key.
The Pacer XD does not yet generate stimulation impulses. First the parameters
Operating mode
Stimulation Frequency
Stimulation Intensity
need to be set and the Pacer must be started.
The PRIMEDIC
Pacer XD is switched off by pressing the Pacer On- / Off-key. Simul-
taneously 3 acoustic warning signals are given.
If the PRIMEDIC
Pacer XD is activated without the Pacer- / Defibrillation cable con-
nected to the socket of the Pacer XD, an acoustic warning is emitted and the message
"Pacer Electrodes Open" appears on the monitor.
If a warning sounds after activating the pacer for each pacemaker impulse, the device
has detected a deviation of ± 15 % from the set intensity. A possible cause of this can
be patient resistance that is too great. If a deviation of ± 30 % is detected, the device
interrupts the pacer process automatically.
The Pacer XD is powered by the battery of the PRIMEDIC
monitor the battery capacity gauge in the monitor during pacing. If the message
"Charge Accu" appears, the battery has to be recharged or exchanged to continue
the stimulation. If the Pacer is used with low battery harm to the patient by device
failures is possible.
If the activated PRIMEDIC™ pacer is operated for longer than 3 minutes, it switches
off automatically.
Pacer XD the following operating order must be con-
Pacer XD.
DefiMonitor XD
Defi-Monitor. Therefore,
21324 / 09.05 / 02 / GB