4.7
4.7.1
4.7.2
22
HeartSave AED / AED-M
Device Self Test
Self Test After Switch On Of HeartSave
The self test will either be activated when switching on the HeartSave or inserting
the energy module into the device and passes immediately the device self test to
check all important functions and signal settings.
If the energy module was changed and the HeartSave has located a failure the auto-
matic self test (FULL) will take place. Follow the instructions of the HeartSave.
Periodical self test
Cycle
SHORT
Daily
MEDIUM
First day of the month
LONG
On July 1st and January
1st of each year
Operating Manual
Test
Software, keyboard, battery, ECG
calibration, clock, internal voltage
supply and HV part at 0 V
Software, keyboard, battery, ECG
calibration, clock, internal voltage
supply and HV part at 300 V
Software, keyboard, battery, ECG
calibration, clock, internal voltage
supply and HV part at 1600 V
20439 / 03.04 / 03 / engl.