H
Intermediate height
The unit is defined in the Distance unit parameter (→  128).
 Setup → Advanced setup → Linearization → Table mode
Linearization type (→  148) = Table
Select editing mode of the linearization table.
• Manual
*
• Semiautomatic
• Clear table
• Sort table
Manual
Meaning of the options
• Manual
The level and the associated linearized value are entered manually for each linearization
point.
• Semiautomatic
The level is measured by the device for each linearization point. The associated
linearized value is entered manually.
• Clear table
Deletes the existing linearization table.
• Sort table
Rearranges the linerization points into an ascending order.
Levelflex FMP53 FOUNDATION Fieldbus
H
0%
A0013264