Endress+Hauser Levelflex FMP53 FOUNDATION Fieldbus Operating Instructions Manual page 87

Guided wave radar
Hide thumbs Also See for Levelflex FMP53 FOUNDATION Fieldbus:
Table of Contents

Advertisement

Levelflex FMP53 FOUNDATION Fieldbus
Endress+Hauser
Functions of the parameter options
Options
Description
33097: Execute backup
A backup copy of the current device configuration in the HistoROM is saved to the
display module of the device. The backup copy comprises the transmitter data of
the device.
33057: Restore
The last backup copy of the device configuration is copied from the display module
to the HistoROM of the device. The backup copy comprises the transmitter data of
the device.
33838: Duplicate
The transmitter configuration from another device is duplicated to the device using
the display module.
265: Compare
The device configuration saved in the display module is compared to the current
device configuration of the HistoROM.
32848: Clear backup data
The backup copy of the device configuration is deleted from the display module of
the device.
HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
For devices with FOUNDATION Fieldbus communication, the PD Tag parameter is
also transmitted when duplicating the parameter configuration. If required change PD
Tag to the required value after duplicating the set.
Commissioning (block-based operation)
87

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Levelflex FMP53 FOUNDATION Fieldbus and is the answer not in the manual?

Questions and answers

Table of Contents