WARNING

4.2. Testing the Precision Platform operation

4-2
and document the test. The main contact numbers for Radiological
Operations are
First level of support:
• The Americas, Asia Pacific and India are supported directly by
Phoenix at 602-313-3330.
• Europe, Middle East and Africa are supported by Waterford at
+ 353 (0) 51 372 151.
Second level of support (world-wide) is Phoenix at 602-313-3330. Phoenix
also has a toll-free number for the U.S. and Canada at 866-811-0312.
Information on licensing and regulatory requirements, radioactive sources
and safety is available in the Honeywell Radiation Safety Training Manual
(p/n 6510020199).
Failure to follow radiation safety procedures MAY cause exposure
:
to hazardous levels of radiation.
Once the Precision Platform has been installed and set up, test the
subsystems listed (in the order shown), and make any needed adjustments
as described in the sections:
• Drive motor torque
• Main drive belt tension
• Head alignment
• Internal head temperatures
• Sheet guide temperatures
Testing of the Precision Platform includes verifying the scanner scans and
scans smoothly without the drive motor torque being set too high – the
scanner heads should be able to be stopped by hand. Testing also includes
10/4/07
System Testing and Verification
P/N 6510020208