Print Records Manually - Beckman Coulter MET ONE 3411 User Manual

Table of Contents

Advertisement

4. To show the average of the values in the printout when the samples
are taken and the data is printed, select Print Averages.
Note: The fourth value in the printout is the average of the three previous
values.
5. To round values to the nearest whole number, select Round
Averages.
6. To repeat the header data on all samples, select Repeat Sample
Headers. When not selected, the headers print only with the first
sample of a series when the cycles are set to > 1.
7. Push ENTER to confirm.

Print records manually

The buffer holds 5000 records maximum. The Print Center can print the
entire buffer or the average of count cycles obtained at each sample
point. To print records manually:
1. On the counter Navigation screen, push PRINTER.
2. On the Print Center screen, select the print option for the data. In the
Print Report field, select the regulatory reports that will be printed.
• Push AVERAGE to print the average of each size channel for the
last sample measurement. Refer to
• Push BUFFER to print all of the records in the buffer. Refer to
Figure
7.
3. The data will begin to print.
• To cancel the print job, push CANCEL PRINT.
• To return to Counter Navigation, push RETURN.
Figure
6.
Figure 6 Averages report
English 35