Configure and enable the FTP function

With firmware V4.08, the particle counter can be configured to transmit
data via Ethernet (wired or wireless when the wireless option is installed)
to an FTP server. This particle counter can connect to servers that use
FTP or FTPS (Explicit TLS/SSL).
1. For the user account that will be used in the particle counter, give file
access on the FTP server: Read, Write and Delete.
2. For the user account that will be used in the particle counter, give
directory access on the FTP server: Create, Delete, List and Add
Subdirectories.
Note: The Create, Delete and Add Subdirectories rights are only necessary if
the particle counter will be configured to make subdirectories. These rights are
not necessary when the particle counter is configured to use the existing
directories.
3. On the Counter Navigation screen, push EXPORT.
4. Select Configure FTP.
5. Select Network Server Enabled.
6. In the Host/IP Addr field, enter the host name or the IP address of
the computer where the FTP server is installed.
7. In the Port field, enter the port number on which the FTP server
listens.
8. In the Protocol field, select the protocol the FTP server uses (FTP or
FTPS (TLS/SSL)).
9. Optional: In the Initial Folder field, enter the initial folder where files
will be kept. Leave blank to save files to the root.
10. Select the Logon tab.
11. In the Authentication field, select Anonymous or Normal.
12. If Normal was selected, enter a user name and password to use for
authentication with the FTP server.
13. Push Test Connection. When the connection is successful, "Test
Connection Succeeded" shows.
Note: If the connection was not successful, an error message shows with the
reason.
38 English

Data transfer to the OPC server

Only administrators, factory service personnel and operators with User
Upload permission can initiate data upload to the OPC server.
There are three settings that control the data transfer to the OPC server:
User initiated upload, user-initiated download and automatic download.

User-initiated data upload

Use this option to control when data is sent to the server.
1. Go to the Network screen.
2. Push
to initiate the data transfer.

User-initiated data download

User-initiated download is done from the OPC server. Use this option to
verify the server connection and start the data download.
To start the data download, push DOWNLOAD NOW. All other actions
are done by the server.

Automatic data download

Automatic download is configured on the OPC server. For applications
that require alarm reasons, automatic download must be disabled.

Calibration

The instrument cannot be calibrated by the user. Contact the
manufacturer for instrument calibration.

Maintenance

W A R N I N G
Multiple hazards. Do not disassemble the instrument for maintenance.
If the internal components must be cleaned or repaired, contact the
manufacturer.