Clear Manually Assigned Racks - Beckman Coulter UniCel DxC 600 Daily Start Up

Synchron clinical system
Hide thumbs Also See for UniCel DxC 600:
Table of Contents

Advertisement

Clearing Racks does not clear programming for Sample ID's; it only
clears the assignment to a rack/position. Your Closed Tube Tracking
Database tracks which samples have been pierced by Sample ID.
You may get a message informing you that some racks are In-Progress,
Removed, Requested for calibration, or Reserved for Cal. This feature
protects you from inadvertently clearing a rack that is needed for one of
these conditions.
UniCel DxC, Customer, Daily Start Up Job Aid
Rev. F (September 2013)

Clear Manually Assigned Racks

Select Samples from the menu bar
Select Clear F7
Select the Rack(s) field. Type the range of
rack numbers(s) to be cleared
Note: Enter 1-999 to clear all racks
Select Enter
from the keyboard
Select OK
Select OK
to confirm
Notes about Clearing Racks