LABORIE MMS Solar Blue User Manual page 85

Table of Contents

Advertisement

The moment the profile starts (just before the urethral pressure increases
because the catheter reaches the beginning of the urethra), the transducer
measuring the urethral pressure Pura is still in the bladder; hence, this is the
resting bladder pressure.
Calculated: Resting bladder pressure = [F].
Maximum Urethral Closure Pressure (MUCP)
Calculation method
The Maximum Urethral Closure Pressure (MUCP) can be calculated in two
ways:
• MUCP = Max P
• MUCP = Max (P
In the analysis program, you can define how to calculate the Maximum
Closure Pressure. From the menu, choose Settings > UPP settings.
Figure 7.4 MUCP calculation method
MUCP = Max P
ura
MUCP is the difference between the maximum urethral pressure and the
resting bladder pressure.
Max P
is the maximum urethral pressure (P
ura
marker and the Profile end marker. P
This is the pressure at the position of the Profile start marker on the urethral
channel (P
). The moment the profile starts (just before the urethral pressure
ura
increases because the catheter reaches the beginning of the urethra), the
transducer measuring the urethral pressure (P
this is the resting bladder pressure.
MUCP = Max (P
ura
Urethral Pressure Profile (UPP)
– P
(ura)
ura
ves
– P
) default
ura
ves
– P
(ura)
ves
– P
)
ves
Solar Blue User Manual
) between the Profile start
ura
(ura) is the resting bladder pressure.
ves
) is still in the bladder, hence
ura
85

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents