Rest profile (use the catheter puller)
Press the Start Puller button to start the puller. An increase in urethral
pressure (P
When P
ura
inserted automatically (backwards in time). The position of the marker
will be determined by the software and is based on P
When P
ura
inserted automatically. Also, the Maximum Urethral Closure Pressure
marker (MUCP) will be automatically inserted on the highest pressure.
Press the Stop puller button to stop the puller. The puller automatically
returns to its starting position. Position both sensors (P
the bladder.
Stress profile (use the catheter puller)
Press the Start Puller button to start the puller. Ask the patient to
cough every two seconds. The peaks caused by coughing are used to
calculate transmissions. An increase in urethral pressure (P
indicates that the urethral sensor enters the urethra. When P
increases > 10 cmH2O, a Profile start marker will be inserted
automatically (backwards in time). The position of the marker will be
determined by the software and is based on P
decreases to a certain value, a Profile end marker will be inserted
automatically. Also, the Maximum Urethra Closure Pressure marker
(MUCP) will be automatically inserted on the highest pressure.
Press the Stop puller button to stop the puller. The puller automatically
returns to its starting position. Position both sensors (P
the bladder.
Repeat the measurements if needed. At different filling volumes, you may stop
the pump to perform rest and stress profiles.
Press the Stop investigation button.
Urethral Pressure Profile (UPP)
) indicates that the urethral sensor enters the urethra.
ura
increases > 10 cmH2O, a Profile start marker will be
decreases to a certain value, a Profile end marker will be
Solar Blue User Manual
at the start.
ura
and P
ves
)
ura
ura
at the start. When P
ura
and P
ves
) in
ura
ura
) in
ura
81