Figure 9.11 Sensation settings
During the sensation test, different sensation markers can be placed. The
following markers can be enabled in the protocol: no sensation, first sensation,
first urge, modest urge, intense urge, maximum tolerated volume, and pain.
The setting Display balloon markers can be (de)selected.
Balloon filling settings
On the Filling tab, you can set the filling method. For the Syringe filling
method, you can enter the volume increment (e.g., 10 ml) or edit a complete
list (e.g., 10-10-10-20-20). You can also select Ask balloon volume. After
inserting a sensation marker during the test, the software will ask to enter the
balloon volume.
For filling the balloon with the pump (water) you can enter the pump speed
(default 50 ml/min) on the Filling tab.