LABORIE MMS Solar Blue User Manual page 35

Table of Contents

Advertisement

It is always possible to connect the Bluetooth flowmeter to the Solar Blue
module by using a digital MMS-bus cable. All of the following three conditions
must be met:
• The Solar Blue module needs to communicate with the computer via USB.
• The Solar Blue module must be powered by the external power supply.
• The flowsensor must be connected to one of the digital MMS-bus ports of
the Solar Blue module. (Note: it is not possible to connect the power
adapter to the flowsensor as it is powered by the Solar Blue external power
supply.)
Status LED flowsensor
The following table describes what the LED colors and status indicate:
LED
Orange on
Green on
Green blinking
(short periods off)
Green blinking
(long periods off)
Red blinking
Off
Investigation Procedure
Indication
IPL error. Please contact your MMS representative.
Digital cable is connected.
Bluetooth communication flowsensor powered by power
adapter.
Bluetooth communication flowsensor powered by
batteries.
Low battery message, place new batteries in the battery
compartment.
No or empty batteries. Place new batteries in the battery
compartment.
Power adapter (optional) is not connected. Reconnect
the power adapter.
Bluetooth flowsensor in low power mode.
Bluetooth connection is lost. In case you are using the
Bluetooth dongle, check whether the dongle is missing
or broken.
Solar Blue User Manual
35