Thermo Scientific
CAUTION Because intense UV light can damage your eyes, the detector automatically
turns off the lamps when the lamp cover is removed.
CAUTION Because the lamps run hot, turn off the power and allow sufficient time
(approximately 30 minutes) for the lamps to cool before you remove the lamp cover.
3. Using the 1/8 in. ball-end hex screwdriver (supplied), unscrew the captive retaining screw
on the lamp cover (see
4. Pull the cover straight back to expose the lamp assemblies.
5. Gently unplug the tungsten lamp lead from the detector (see
Take care not to twist the connector.
Figure 53. Tungsten lamp assembly
Tungsten lamp
connector
6. Loosen and remove the thumbscrew that holds the lamp assembly in place.
7. Pull the assembly straight out and remove it from the detector.
Figure 48
on
page
49).
Thumbscrew
Tungsten lamp
assembly
3
Routine Maintenance
Lamp Maintenance
Figure
51).
Accela UV/Vis Detector Hardware Manual
53