2
Installation
Installing the Standard Analytical Flow Cell
22
Accela UV/Vis Detector Hardware Manual
Figure 26. Tubing clamp
Tubing clamp
b. Using the 3/32 in. ball-end hex screwdriver (supplied), remove the two photodiode
assembly mounting screws and set them aside.
c. Carefully pull the photodiode mount toward you and off the flow cell assembly.
Rest the photodiode mount out of the way inside the detector housing
Figure 27. Photodiode mount removed
d. Using the 9/64 in. ball-end hex screwdriver (supplied), remove the two screws
securing the LightPipe flow cell mounting assembly to the detector (see
5. Store the LightPipe flow cell mounting assembly in a secure place for possible future use.
The mounting area is now ready for the standard analytical flow cell (see
(Figure
27).
Flow cell mounting
assembly screws
Figure
27).
Figure
28).
Thermo Scientific