Thermo Scientific Accela UV/Vis Plus Hardware Manual page 44

Table of Contents

Advertisement

2
Installation
Making the Back Panel Connections
Figure 35. System interconnect cable for an Accela LC system with an Accela Autosampler
To
MS detector
28
Accela UV/Vis Detector Hardware Manual
M/S
To
detector
To
autosampler
To
pump
To make the contact closure connections for a system with an Accela Autosampler
1. To connect the UV/Vis detector, plug one of the three connectors labeled DETECTOR
into the left, 8-pin receptacle located on the back panel of the UV/Vis detector as shown
in
Figure 9
on
page
6.
Note If your LC system contains more than one detector, contact your Thermo
Fisher Scientific field service engineer for instructions on how to interconnect the run
signals for the Accela LC modules.
2. To connect an autosampler, plug the unlabeled connector into the left, 8-pin receptacle
on the back panel of the autosampler as shown in
3. To connect an Accela pump (all models), plug a connector labeled PUMP into the 8-pin
receptacle on the back panel of the pump.
A/S
Figure
36.
To
detector
To
detector
To
pump
Thermo Scientific