Thermo Scientific Accela UV/Vis Plus Hardware Manual page 35

Table of Contents

Advertisement

Thermo Scientific
Figure 21. Retaining bar in the alignment slot on the 10 mm LightPipe flow cell
d. Make sure that the outlet tubing (blue) on the 10 mm LightPipe flow cell lies in a slot
on the right side of the retaining block.
Figure 22. Retaining bar in the alignment slot on the 50 mm LightPipe flow cell
6. Reinstall the flow cell retaining block as follows:
a. Align the retaining block with the notch in the base of the assembly (see
Figure 23. Retaining block in place
Installing the LightPipe Flow Cell
Notch
Accela UV/Vis Detector Hardware Manual
2
Installation
Figure
23).
19