Figure 28. Empty flow cell mounting area
6. Install the standard analytical flow cell as follows:
a. Position the standard analytical flow cell assembly against the flow cell mounting
area.
b. Using the 9/64 in. ball-end hex screwdriver (supplied) and the two screws removed
from the LightPipe flow cell mounting assembly, secure the standard analytical flow
cell assembly to the detector (see
Figure 29. Standard analytical flow cell mounted to the detector
c. Position the photodiode assembly with the mounting holes aligned with the two
standoffs on the flow cell assembly (see
Figure
29).
Standoffs
Figure
29), and using a 3/32 in. ball-end hex
Accela UV/Vis Detector Hardware Manual
2
Installation
Installing the Standard Analytical Flow Cell
Flow cell mounting
alignment pins
Flow cell mounting
screws
23