Light
25028
DMD1 temperature_Too high.
Engine
Light
25029
DMD2 temperature_Low.
Engine
Light
25030
DMD2 temperature_Too low.
Engine
Light
25031
DMD2 temperature_High.
Engine
Light
25032
DMD2 temperature_Too high.
Engine
Light
Diffusor motor
25033
Source
temperature_Low.
Light
Diffusor motor
25034
Source
temperature_Too low.
Supra-5000 Digital Cinema and Secured Media Projection System
User Manual
1. The projection room temperature
is higher than 35℃.
2. The air inlet or air outlet of the
projector is blocked.
3. The filter of the projector is
clogged.
4. The system fan is faulty.
5. If the temperature value does not
change, the NTC is faulty.
1. The projection room temperature
is below 0℃.
2. If the temperature value does not
change, the NTC is faulty.
1. The DMD NTC may be open-
circuited or damaged.
2. The projection room temperature
is below –5℃.
1. The projection room temperature
is higher than 35℃.
2. The air inlet or air outlet of the
projector is blocked.
3. The filter of the projector is
clogged.
4. The system fan is faulty.
5. If the temperature value does not
change, the NTC is faulty.
1. The projection room temperature
is higher than 35℃.
2. The air inlet or air outlet of the
projector is blocked.
3. The filter of the projector is
clogged.
4. The system fan is faulty.
5. If the temperature value does not
change, the NTC is faulty.
1. The projection room temperature
is below 0℃.
2. If the temperature value does not
change, the NTC is faulty.
1. The diffusor motor NTC may be
open-circuited or damaged.
2. The projection room temperature
is below –5℃.
- 184 -
1. Check the projection room temperature.
2. Check whether there are obstacles at the air inlet or air
outlet.
3. Check and replace the air inlet filter.
4. Check and replace the faulty system fan.
5. Replace the NTC (on the light engine) of the DMD1
radiator.
1. Check the projection room temperature and ensure
that the indoor temperature is higher than 0℃.
2. Replace the NTC (on the light engine) of the DMD2
radiator.
1. Replace the DMD NTC.
2. Ensure that the projection room temperature is higher
than –5℃.
1. Check the projection room temperature.
2. Check whether there are obstacles at the air inlet or air
outlet.
3. Check and replace the air inlet filter.
4. Check and replace the faulty system fan.
5. Replace the NTC (on the light engine) of the DMD2
radiator.
1. Check the projection room temperature.
2. Check whether there are obstacles at the air inlet or air
outlet.
3. Check and replace the air inlet filter.
4. Check and replace the faulty system fan.
5. Replace the NTC (on the light engine) of the DMD2
radiator.
1. Check the projection room temperature and ensure
that the indoor temperature is higher than 0℃.
2. Replace the diffusor motor sensing board (on the light
source).
1. Replace the diffusor motor NTC.
2. Ensure that the projection room temperature is higher
than –5℃.