GDC Espedeo Supra Supra-5000 User Manual page 183

Digital cinema and secured media projection system
Table of Contents

Advertisement

Light
25020
PCS temperature_Too high.
Source
Light
25021
Engine
temperature_Low.
Light
Optical prism temperature_Too
25022
Engine
Light
25023
Engine
temperature_High.
Light
Optical prism temperature_Too
25024
Engine
Light
25025
DMD1 temperature_Low.
Engine
Light
25026
DMD1 temperature_Too low.
Engine
Light
25027
DMD1 temperature_High.
Engine
Supra-5000 Digital Cinema and Secured Media Projection System
User Manual
1. The projection room temperature
is higher than 35℃.
2. The air inlet or air outlet of the
projector is blocked.
3. The filter of the projector is
clogged.
4. The PCS fan is faulty.
1. The projection room temperature
Optical prism
is low.
2. If the temperature value does not
change, the NTC is faulty.
1. The optical prism NTC may be
open-circuited or damaged.
low.
2. The projection room temperature
is below –5℃.
1. The projection room temperature
is higher than 35℃.
2. The air inlet or air outlet of the
projector is blocked.
Optical prism
3. The filter of the projector is
clogged.
4. The system fan is faulty.
5. If the temperature value does not
change, the NTC is faulty.
1. The projection room temperature
is higher than 35℃.
2. The air inlet or air outlet of the
projector is blocked.
3. The filter of the projector is
high.
clogged.
4. The system fan is faulty.
5. If the temperature value does not
change, the NTC is faulty.
1. The projection room temperature
is below 0℃.
2. If the temperature value does not
change, the NTC is faulty.
1. The DMD NTC may be open-
circuited or damaged.
2. The projection room temperature
is below –5℃.
1. The projection room temperature
is higher than 35℃.
2. The air inlet or air outlet of the
projector is blocked.
3. The filter of the projector is
clogged.
4. The system fan is faulty.
5. If the temperature value does not
change, the NTC is faulty.
1. Check the projection room temperature.
2. Check whether there are obstacles at the air inlet or air
outlet.
3. Check and replace the air inlet filter.
4. If the PCS fan is determined to be faulty; the light
engine module may need to be returned to the factory for
processing. Please contact customer service personnel.
1. Check the projection room temperature and ensure
that the indoor temperature is higher than 0℃.
2. Replace the optical prism NTC.
1. Replace the optical prism NTC.
2. Ensure that the projection room temperature is higher
than –5℃.
1. Check the projection room temperature.
2. Check whether there are obstacles at the air inlet or air
outlet.
3. Check and replace the air inlet filter.
4. Check and replace the faulty system fan.
5. Replace the optical prism NTC.
1. Check the projection room temperature.
2. Check whether there are obstacles at the air inlet or air
outlet.
3. Check and replace the air inlet filter.
4. Check and replace the faulty system fan.
5. Replace the optical prism NTC.
1. Check the projection room temperature and ensure
that the indoor temperature is higher than 0℃.
2. Replace the NTC (on the light engine) of the DMD1
radiator.
1. Replace the DMD NTC.
2. Ensure that the projection room temperature is higher
than –5℃.
1. Check the projection room temperature.
2. Check whether there are obstacles at the air inlet or air
outlet.
3. Check and replace the air inlet filter.
4. Check and replace the faulty system fan.
5. Replace the NTC (on the light engine) of the DMD1
radiator.
- 183 -