1.4 Electromagnetic Compatibility (EMC)

1.4.1 Electromagnetic Compatibility (EMC) Related Standard

202B1262871D
To avoid electromagnetic interference in the operational environment, read the following
precautions and properly handle this product and other devices in the vicinity.
This product has been tested and confirmed to comply with the limits for medical devices
defined in EN 60601-1-2:2007. These limits are designed to provide reasonable protection
against harmful electromagnetic interference in a typical installation at medical facilities.
However, there is a possibility that this product may cause harmful electromagnetic
interference to other devices in the vicinity, even if it is used according to the processor
instructions. Also, there is no guarantee that interference will not occur in a particular
installation. If this product does cause harmful electromagnetic interference to other
devices, that can be determined by turning the processor off and on, we recommend that
you may try to correct the interference by one or more of the following measures:
• Change the orientation or position of any affected device.
• Increase the spacing between devices.
• Consult the manufacturer or dealer of other devices.
If the problem cannot be solved with the above measures, stop using this product and
consult the manufacturer or your local FUJIFILM dealer for help.
Do not place any objects that emit strong electromagnetic waves near
this product. Otherwise, malfunction of this product may occur.
Noise may appear on the monitor of this product due to the effect of
electromagnetic waves. In this case, turn off the device emitting the
electromagnetic waves or move the device away from this product.
Use the cable specified in the Installation Manual of the processor
for this product. Use of other cables may cause an increase in
electromagnetic emission or decrease in electromagnetic immunity of
this product.