3.9 Attaching and Inspecting Balloon

(1) Feed sterile water into the balloon water feed channel by
fully pressing the air/water valve, and discharge air inside
of the balloon water feed channel.
202B1262871D
Do not use this product with an endoscopic CO
bubbles may be generated in the sterile water inside the water tank
and enter the balloon channel. This could prevent water discharge from
the balloon and disable deflation of the balloon.
Use a disinfected (or sterilized) balloon attachment tool. There is a risk
of infection.
Do not use a balloon on patients allergic to latex. There is a risk of an
anaphylactic reaction.
Do not feed more than 5 ml of water into the balloon. There is a risk of
rupturing the balloon.
[Note] It takes approximately 3 seconds to feed in 5 ml of water. Check the amount of
water to be fed before use.
[Note] If water cannot be discharged or little water is discharged from the balloon, refer to
"Chapter 5 Troubleshooting".