FujiFilm EG-580UR Operation Manual page 87

Ultrasonic endoscope
Table of Contents

Advertisement

Problem
Live image is not
displayed after image
freezing is canceled
during a procedure.
The endoscopic
image is discolored
during a procedure.
[Note] To reset the processor and the light source, turn them off, and wait for at least 5 seconds. Turn on the processor and
the light source again, and then light the lamp by pressing the Lamp button.
[Note] To reset the ultrasonic processor, turn it off, and wait for at least 5 seconds. Turn on the ultrasonic processor again.
202B1262871D
Cause
The system is malfunctioning possibly due to
static charge.
1) The imaging section is damaged.
2) The system is malfunctioning possibly due
to static charge.
3) The video signal cable has shorted out or
broken.
Chapter 5 Troubleshooting
Remedy
If the problem occurred during treatment,
stop treatment and remove the endotherapy
device from the ultrasonic endoscope. Then,
[Note]
reset
the processor and the light source. If
the problem occurred during an examination,
[Note]
reset
the processor and the light source.
If an appropriate image does not appear
even after resetting, turn them off, straighten
the bending section to unlock, release the
angulation knobs, withdraw the ultrasonic
endoscope slowly from the patient.
1) 2) 3)
If the problem occurred during treatment,
stop treatment and remove the endotherapy
device from the ultrasonic endoscope.
[Note]
Then, reset
the processor and the light
source. If the problem occurred during an
[Note]
examination, reset
the light source. If an appropriate image
does not appear even after resetting and it
is impossible to continue the examination,
turn them off, straighten the bending section
to unlock, release the angulation knobs,
withdraw the ultrasonic endoscope slowly
from the patient.
the processor and
87