Cautions Relating to Electromagnetic Compatibility
Cautions Relating to
Electromagnetic Compatibility

Electromagnetic Compatibility (EMC)

This equipment complies with the EMC standards applicable to medical
devices, in accordance with the requirements of IEC60601-1-2 :2001 +
 A1 2004.
However, there is no guarantee that harmful electromagnetic
interference will not occur in any installation environment.
This equipment generates, and can radiate radio frequency energy. In a
particular environment, e.g., the equipment is not installed and used in
accordance with the instructions or any device that is not compliant with
the EMC Standard is used in a surrounding area, harmful interference
may result, causing the equipment or surrounding device to malfunction.
If this equipment causes harmful interference to other equipment or
receives harmful interference from other equipment, you are
encouraged to try to correct the interference by one or more of the
following measures:
Reorient or relocate the equipment.
Increase the separation between the equipment and the affected
equipment.
Connect the equipment into an outlet on a circuit different from that to
which the affected equipment is connected.
If the problem persists, discontinue the use of the equipment and
consult the manufacturer of the affected equipment or FUJIFILM
JAPAN.
Further information for IEC60601-1-2:2001
(1) Medical electrical equipment needs special precautions regarding
EMC and needs to be installed and put into service according to the
EMC information provided in the accompanying documents.
(2) Portable and mobile RF communications equipment can affect
medical electrical equipment.
(3) Information of all cables and their maximum lengths and accessories
that may affect EMC.
Name
N et wo rk ca bl e
C at5 e o r h ig he r, UTP -ty pe st ra ig ht ca bl e
(4) The use of accessories, transducers and cables other than those
specied, with the exception of transducers and cables sold by
[FUJIFILM Corporation] as replacement parts for internal
components, may result in increased emissions or decreased
immunity of the [CR-IR346CL/348CL].
(5) The [CR-IR346CL/348CL] should not be used adjacent to or stacked
with other equipment.
If adjacent or stacked use is necessary, the [CR-IR346CL/348CL]
should be observed to verify normal operation in the conguration in
which it will be used.
(6) Test items (Tables 1 to 4)
Table 1
Guidance and manufacturer's declaration - electromagnetic emissions
The CR-IR346CL/348CL is intended for use in the electromagnetic
environment specied below. The customer or the user of the CR-
IR346CL/348CL should assure that it is used in such an environment.
Emissions test
Compliance Electromagnetic environment - guidance
RF emissions
Group 1
CISPR 11
Class A
RF emissions
Group 1
CISPR 11
Class A
Harmonic emissions
Class D
IEC 61000-3-2
Voltage uctuations/
Not
icker emissions
complied.
IEC 61000-3-3
Specication
The CR-IR346CL/348CL uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
The CR-IR346CL/348CL is suitable for
use in all establishments other than
domestic and those directly connected to
the public low-voltage power supply
network that supplies buildings used for
domestic purposes.