1.4 Electromagnetic Compatibility (EMC)

This equipment has been tested and found to comply with the limits for medical devices to the IEC
60601-1-2:2001+Amd1:2004/EN 60601-1-2:2001+Amd1:2006.
These limits are designed to provide reasonable protection against harmful interference in a typical
medical installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference to other devices in the
vicinity.
However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to other devices, which can be determined by
tuning the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving device.
• Increase the separation between the equipment.
• Connect the equipment into an outlet on a circuit different from that to which the other device(s)
are connected.
Consult the manufacturer or field service technician for help.
Further information for IEC 60601-1-2:2001/EN 60601-1-2:2001
• Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in the accompanying documents.
• Portable and mobile RF communications equipment can affect medical electrical equipment.
• Information regarding the cable affecting EMC is as follows.
Name
Network Cable
• The use of accessories, transducers and cables other than those specified, with the exception
of transducers and cables sold by FUJIFILM Corporation as replacement parts for internal
components, may result in increased emissions or decreased immunity of the FCR PRIMA T2
(CR-IR 392).
• The FCR PRIMA T2 (CR-IR 392) should not be used adjacent to or stacked with other
equipment.
If adjacent or stacked use is necessary, the FCR PRIMA T2 (CR-IR 392) should be observed to
verify normal operation in the configuration in which it will be used.
General Specification
Cat 5 or more, UTP type and straight cable.