8. INSTALLATION

Overview
This section provides detailed installation instructions for the CLOVER A1c
must follow the installation steps correctly to ensure proper installation, operation, and
service.
CAUTION
Do not drop or handle the system roughly. The calibrated optics, electronics orother
internal parts can be disrupted or damaged. Always handle the system with care.
The CLOVER A1c
Self ia s precise system and must be handle carefully.
TM
Place the system where it will not be subjected to extreme temperature variations. Avoid
open windows, direct sunlight, ovens, hot plates, open burners, radiators and dry ice
baths.
Do not place the CLOVER A1c
TM
Self on the same place with a source of vibration
Provide large space to allow free air circulation around the system - 8cm(3 inches) on all
sides.
Unpacking the System
Your CLOVER A1c
Self is delivered in 1 shipping carton.
TM
1.Carefully take out the contents of the shipping carton.
2.Please check physical damage on the carton and system
3.If the system is damaged, please contact to your representative in your area.
4. Make sure all items are included with your system, and keep them for future use.
Installing Connections
3
4
5
1
2
Connect the System Power
1. Ensure that the system power button is in the off position.
2.Connect the power code of the system to appropiately grounded AC electrical outlet.
CLOVER A1c Self • 10 •
TM
Self. You
System Connections:
1. Power Button
2. DC 9 V
3. USB port
4. Barcode Scanner port
5. Thermal Printer port (Serial port RS 232)

9. OPERATION

9.1 Power On
Connect the DC 9V adapter to the power port on the back side of the analyzer.
On/off
9V
USB
Barcode
Make sure that the cover is closed before connecting the adapter.
After the system is properly installed, you can start the CLOVER A1c
To start the CLOVER A1c
TM
Self, turn the power
switch to the 'ON' position.
08/03/30
AM10:03
HbA1c
Glucose
If the cover is open, the icon of "Close the cover" will
be displayed.
Close the cover, and the warming up will start.
08/03/30
AM10:04
HbA1c
Glucose
If the cartridge is inserted, the icon of "Remove the
Remove
cartridge" will be displayed.
Take out the cartridge and close the cover.
Warming up will start.
CLOVER A1c Self • 11 •
Printer
TM
Self.