Infopia GluNEO Lite Instructions For Use Manual

Infopia GluNEO Lite Instructions For Use Manual

Blood glucose monitoring system for self testing

Advertisement

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GluNEO Lite and is the answer not in the manual?

Questions and answers

Александр
April 27, 2025

Аксессуары какой фирмы можно использовать

1 comments:
Mr. Anderson
May 14, 2025

The context does not specify any brand accessories that are compatible with Infopia Lite. Therefore, compatibility information cannot be determined.

This answer is automatically generated

Santhna
February 5, 2025

Code 126 on the machine when the strip is inserted

SHARIQ AFTAB KHAN
February 2, 2025

How to set time and date and connect with blue tooth

User image 679efc28710a8

Subscribe to Our Youtube Channel

Summary of Contents for Infopia GluNEO Lite

  • Page 2 Blood Glucose Monitoring System for self testing...
  • Page 3: Important Information

    ® Dear GluNEO Lite Blood Glucose Monitoring System Owner, ® Thank you for choosing the GluNEO Lite Blood Glucose Monitoring System. This manual contains everything you need to know about your new glucometer and how it works. Please take a moment to read the instructions carefully. We understand that self-testing of blood glucose level provides a way to control your ®...
  • Page 4: Table Of Contents

    Table of Contents Test Principle 1. Learning the System ® GluNEO Lite blood glucose test strip is the bio-sensor which is composed of the Intended use of blood glucose test meter--------------------------------------------------- 6 FAD-dependent glucose dehydrogenase(enzyme). FAD-GDH(enzyme) quantitatively ® GluNEO Lite kit contents ---------------------------------------------------------------------- 7 reacts with the glucose in the whole blood and the mediator.
  • Page 5: Learning The System

    ® GluNEO Lite Kit Contents Learning the System Intended use ® ® The GluNEO Lite test meter is used with GluNEO Lite test strip for the measurement of glucose in fresh whole blood samples(Capillary and/or venous). It should not be used for the diagnosis or screening of diabetes or for the testing of new born babies.
  • Page 6: Gluneo ® Lite Meter Display

    ® ® GluNEO Lite Meter Display GluNEO Lite Meter Test Strip Port: Low Battery Warning Symbol ® Insert the GluNEO Lite test strip Alarm Symbol Troubleshooting Symbol Display Screen: Displays your test results, symbols, and messages. Delete Results Power/Enter Button: DATE Power the meter on/off, prompts memory data and acts as Average Test Result System...
  • Page 7: Gluneo ® Lite Test Strip

    ® GluNEO Lite Test Strip Before Testing Test Strip Port WARNING To reduce the chance of infection: 1. Lancing device and lancets should NOT be shared with others. 2. Always use a new lancet and a new blood glucose test strip. 3.
  • Page 8: Setting Your Meter

    Setting Your Meter ® The GluNEO Lite Meter has a wide variety of functions to choose. From the setup mode you can turn activity/meal flags, set the date/time, designate three unique averages, and set up to five daily alarms. Temperature Unit Number of days for Alarm on/off After pressing and...
  • Page 9: Auto-Coding Function

    Performing a Glucose Control Solution Test Auto-coding Function ® GluNEO Lite control solution contains a known amount of glucose that reacts with ® GluNEO Lite test strip. By comparing your control solution test results with the expected range printed on the test strip vial, you are able to check that the meter and test strips are working properly, and that you are performing the test correctly.
  • Page 10 1.Check the expiration date before performing a control For example only solution test. Do not use if expired. Please note the If you are using the normal control solution, according to this particular vial of strips, your expiration date marked on the control solution bottle. meter should show a number between 80 and 120.
  • Page 11: Performing Your Test

    Follow-up Action Performing Your Test 1. Check your meter and test strips with the glucose control solution. Testing Blood - Preparation 2. Perform your test again. Lancing device is a convenient medical tool for collecting capillary blood samples for glucose monitoring or other tests which require one or two drops of blood. Being a more advanced lancing device, it provides ultimate safety and comfort for obtaining CAUTION blood samples.
  • Page 12: Performing A Blood Test With Your Fingertip

    COMPONENTS Performing a blood test with your fingertip Clear Tip Step 1 Ejection System Sliding Barrel Firmly insert the test strip into the meter test port in the direction of arrow on test strip (arrow up). Please do not insert the glucose test strip upside down.
  • Page 13 Step 2 4. Put the lancing tip back and twist it tightly. To obtain a suitable blood samples • Before obtaining a blood sample, wash your hands with warm, clean water and soap. Dry hands completely before testing. WARNING 1. All parts of the kit are considered bio hazardous and can potentially transmit infectious disease, even after you have conducted cleaning and disinfection measures.
  • Page 14 Step 3 Step 4 • Applying Blood • Correct • After the beeping sound, test will begin automatically, and your results will appear in 5 seconds. It should begin counting down from 5 to 1 second on the LCD display window. LCD Completely Filled Correct window will display the result...
  • Page 15: Test Strip Ejector Function

    Test Strip Ejector Function Reviewing Your Results Disposing of your test strip ® The GluNEO Lite meter stores up to 365 test results in its built-in memory, along with the average sugar level for the number of days, which you had preset the glucose test meter to 1.
  • Page 16: Reviewing Your Results With Activated User Id

    Reviewing your results with activated user ID Alternative Site Test Press the button once and the User ID option will appear the LCD screen. Use ▲ or ▼ buttons to scroll and choose the correct Upper Arm user ID, then press the button.
  • Page 17: Lancing Device Clear Tip Is Used For Alternate Site Testing

    Ventral palm/Dorsal hand To ensure accurate results when lancing your arm (forearm or upper arm), or palm (ventral palm or dorsal Choose a fleshy area on the palm, below your thumb or hand), clean the test site with soap and water. pinky finger.
  • Page 18: Caring For Your Gluneo ® Lite System Cleaning Your Meter And Maintenance

    ® Caring for Your GluNEO Lite System 5. Continue to hold the lancing device and gradually increase pressure for several seconds. Cleaning your meter and maintenance Blood glucose test meter : Your blood glucose test meter does not require special maintenance or cleaning. Avoid 6.
  • Page 19: Storage Of Your Gluneo ® Lite System

    ® Storage of your GluNEO Lite System CAUTION ® 1. Store the GluNEO Lite test strip vials in a cool, dry place. Keep out of direct To prevent the meter and test strips from getting dirt, dust or other contaminants, please sunlight.
  • Page 20: Transferring Your Result Data

    Transferring Your Result Data Battery Installation ® You can transfer test results from the GluNEO Lite meter to a computer. The low battery icon will appear in the upper left corner of the LCD display to alert You will need to download the software from the following web site you when the battery power is running low, which is an indication that a new battery is (http://www.infopia21.com) and purchase the computer cable from your local representative.
  • Page 21: Disposal

    Contact your authorized representative or customer support and electronic equipment. center of Infopia Co.,Ltd., if the problem continues. The Waste Electrical and Electronic Equipment(WEEE) regulation implement provisions of the European Parliament and Council Directive 2012/19/EU aimed to reducing the amount of EEE waste going for final disposal.
  • Page 22 Contact your local representative or customer support at Infopia Co, Ltd., if the problem persists. The ambient tempera- Place the meter at a temperature ture is less than 10℃...
  • Page 23: Inconsistent Or Unexpected Test Results

    Alternatively, you are requested to contact your place of purchase in your country for initial warranty request. Service Information Infopia Co.,Ltd. has trained specialists to help you 24 hours a day, 7 days a week, and 365 days a year. IMPORTANT NOTICE Please confirm with the Infopia Co.,Ltd.
  • Page 24: Specifications

    Symbol Reference Specifications Symbol Description Sample Type Capillary and/or venous whole blood Sample Volume 0.5 ㎕ Consult Instructions for use Test Range 20~600mg/dL (1.1 ~ 33.3 mmol/L) Use By date Reading Time 5 seconds This product fulfills the requirements of Directive Calibration Plasma –...