Infopia CLOVER A1c Self Instructions For Use Manual page 28

Test analyzer
Table of Contents

Advertisement

16. TROUBLESHOOTING
ERROR CODE
DESCRIPTION
The HbA1c test
result is higher
>14%
than 14.0%
(130mmol/mol)
Lo
mg/dl
The blood glucose
test result is lower
ou
than 10mg/dl or
0.6mmol/L
Lo
mmol/L
Hi
mg/dl
The blood glucose
test result is higher
ou
than 600mg/dl or
33.3mmol/L
Hi
mmol/L
The
There is a problem
Analyzer
with the power
does not
adaptor or the
analyzer itself
power on
CLOVER A1c Self • 54 •
CORRECTIVE ACTION
Please use the check cartridge to
validate the analyzer and re-test the
blood sample.
Please double check the accuracy of the
test strip by performing a control solution
test and re-test with blood sample.
Please double check the accuracy of the
test strip by performing control solution
test and re-test with a fresh blood sample.
Unplug the power adaptor and reconnect.
If the problem still persists,
contact your local representative.
17. SPECIFICATION
S a m p l e T y p e
Sample Volume
Test Range
Reading Time
Memory Capacity
Wavelength
Power Required
Dimensions/Weight
Storage Temperature
Operating Temperature
Relative Humidity Range
Option
CLOVER A1c Self • 55 •
Capillary whole blood,
Venous blood with anticoagulant
4
4.0~14.0%
5 minutes
200 test results
415 nm (normal)
DC 9 V-1.5 A
200*200*139(mm) / 1.4kg
10 ~ 35
(50 ~ 95
)
17 ~ 32
(63 ~ 90
)
10% ~ 90%
Glucose Test Module
PC Connection Cable
Thermal Printer
Barcode Scanner