14. OPTIONS
STEP 4. Scan the desired barcode until the
Patient ID is displayed on the LCD.
NOTE
It is important that the labels be printed to the required specifications.
Reading errors may occur if any of the following conditions exist :
- The narrow bar width is too small
- The barcode length is too great
- The height is too small
- The reader is held too far from the label
- The background reflection is too high or low
If the reader is not able to consistently read your labels, apply a test label of the format
being used to a new specimen tube and perform the barcode test.
If the reader is able to read the test label, the quality of your labels may be inadequate.
If the test label cannot be read, the reader itself could be the problem.
Contact with our local representatives for technical assistance.
CLOVER A1c Self • 48 •
14. OPTIONS
14.4 Transfer Test Results to a Computer
Analyzer
STEP 1. Connect the RS-232 cable from the analyzer to the PC while in
stand-by mode.
On/Off
9V
USB
Barcode
STEP 2. Once connected, the analyzer will display 'PC'.
STEP 3. Download the PC application program.
STEP 4. Once the program is downloaded, click the 'Date Transfer' icon
on your PC. The results from the analyzer will be transferred over.
CLOVER A1c Self • 49 •
Computer
Printer