11. TESTING HbA1c
STEP 6.
Insert the reagent pack into the cartridge in the analyzer. The 'close the cover' icon will
be displayed.
08/03/30
NOTE :
Do not force the reagent pack into the system. The reagent pack is
designed to fit only in one way.
STEP 7.
The test starts automatically once the cover is closed.
08/03/30
IMPORTANT :
Do not open the analyzer during the test.
Do not vibrate or move the analyzer during the test.
CLOVER A1c Self • 22 •
AM10:03
HbA1c
Glucose
AM10:03
HbA1c
Glucose
4:30
11. TESTING HbA1c
STEP 8.
The timer will countdown from 5 minutes until the HbA1c% is displayed.
08/03/30
AM10:08
HbA1c
Glucose
5.4
%
ID:
NOTE :
If you see ">14.0%" or "<4.0%" displayed, HbA1c% level may be above
14.0% or below 4.0%, repeat testing. If the second result is again outside
this range, please contact your local customer support and service center.
STEP 9.
After the test is done, open the cover of the analyzer. The display will show the 'cartridge
remove' icon. Take out the cartridge from the analyzer by gently pushing to the left and
pulling out.
08/03/30
AM10:09
HbA1c
Glucose
Remove
CAUTION :
Do not give strong impact to cartridge. Dispose all waste in
accordance with applicable national and /or local regulations.
CLOVER A1c™ Self • 23 •