15. PRODUCT LIST
COMPONENTS
ITEM
CONTENTS
1 Analyzer
CLOVER A1c™ Self
1 Instruction for use
1 Exclusive Power Adapter
Analyzer
5 Fan Filters
1 Daily Check Cartridge
1 Monthly Check Cartridge
Optional Products
>>>
CLOVER A1c™ Self
10 Test cartridges
1 Instruction for use
Test Cartridge
CLOVER A1c™ Self
10 Monthly check cartridge
Check Cartridge
CLOVER A1c™ Self
1 Daily check cartridge
Check Cartridge
CLOVER A1c™ Self
1 Glucose Test Module (3Sec)
Glucose Test Modules
1 Instruction for use
(3Sec)
CLOVER A1c™ Self
1 Glucose Test Module (5Sec)
Glucose Test Modules
1 Instruction for use
(5Sec)
1 Thermal printer
Thermal Printer
1 Instruction for use
1 Barcode scanner
Barcode Scanner
1 Instruction for use
PC Cable
1 PC Cable
CLOVER A1c™ Self
1 Fan fileter
Fan Filter
CLOVER A1c Self • 50 •
16. TROUBLESHOOTING
GENERAL INFORMATION
If an operational or system problem occurs, an error code may display on the system
PRODUCT
screen with an explanation of the problem. This section of the guide lists the various
CODE
errors and messages, along with a description and corrective actions. If the problem
persists, record the error code and contact your local representative for technical
support provider for assistance.
INFH01AS
INFHS01AS
INFHMC01AS
INFHDC01AS
AC-CL0-1A1-02
AC-CL0-1A1-02
AC-CLO-1B0-01
AC-CLO-1C0-01
AC-CLO-1D0-01
AC-CL0-1A2-01
ERROR CODE
DESCRIPTION
Problem with the
Er 1
analyzer system
Check Cartridge
Er 2
error
Problem with
Er 3
reading bar code
of test cartridge
Test Cartridge
Er 4
error
Problem with
Er 5
reagent pack of
the test cartridge
CLOVER A1c Self • 51 •
CORRECTIVE ACTION
Please power switch off and on again. If
the problem persists contact your local
representative.
Check cartridge may be damaged (bar
code label, reagent pack, surface of
cartridge etc.) or alien substance was
inserted into the check cartridge.
Press the
button, remove the test
cartridge and then insert again. If the
'reagent pack' icon is displayed, you can
start the test. If the error persists, contact
your local representative.
Test cartridge may be damaged or has
residue on the outside (coding area).
Please press
button and remove the
cartridge. Insert a new test cartridge and
perform the test again.
The analyzer is unable to recognize the
solutions (within the reagent pack). The
reagent pack may be damaged and leak
reagent solution. Please press
button
and remove the cartridge. Insert a new test
cartridge and perform the test again.