13. CHECKING THE SYSTEM
How to use the Daily check cartridge
1) Open the Cover of CLOVER A1c™ Self Analyzer.
2) Press
for 3 seconds to enter into the Daily check cartridge mode.
3) Insert the Daily check cartridge while 'Daily' and 'CHECK' are
blinking.
NOTE : To leave the Daily check cartridge mode, press
CLOVER A1c Self • 30 •
for 3 seconds.
13. CHECKING THE SYSTEM
4) Close the cover. The test starts automatically.
5) After 1 mimute 'OK' or 'Er 2' will be displayed.
6) After the test is complete. Take out the cartridge.
NOTE : If an errer message is displayed, repeat the test. If is still displays Er 2
message 'Do Not use the analyzer'.
Please contact your local representative for customer support.
CLOVER A1c Self • 31 •
'OK' : When the analyzer is ready to
be used.
'Er 2' : When the analyzer is not ready
to be used.