13. CHECKING THE SYSTEM
4) Gently mix the reagent pack 5 ~ 6 times before applying blood sample.
CAUTION :
Do not mix it too violently. It may make some bubbles.
If you drop the test cartridge and the bubble is arisen in the test
cartridge, please wait until the bubble sinks.
5) Insert the Monthly Check Reagent Pack, when the 'insert check reagent pack'
icon is displayed. Insert it respectively.
NOTE : Do not contaminate sample collecting area(Capillary tip) of reagent pack.
CLOVER A1c Self • 34 •
13. CHECKING THE SYSTEM
6) Close the cover. The test starts automatically.
IMPORTANT :
- Do not open the analyzer during the test.
- Do not vibrate or move the analyzer during the test.
7) After 5 minutes 'OK' or errer message will be displayed.
8) After the test is complete. Take out the cartridge.
08/03/30
AM10:09
HbA1c
Glucose
Remove
NOTE : If an error message is displayed, repeat the check again. If it still displays
error message 'Do Not use the analyzer'.
Please contact your local representative for customer support.