Serial and I/O Cable Installation

1. Loosen the 4 Captive screws securing the rear housing to the front housing assembly and
then loosen a gland connector for the serial cable. Refer to Figure No. 2 for an illustration of
the connector layout.
2. Slip the serial cable through the gland connector and into the enclosure.
3. Remove 3" of the outer insulation jacket then remove 1/4" of insulation from each of the
wires (refer to Figure No. 3).
4. Remove the 9-connector terminal block connector from P3 on the 190 mainboard. Grasp the
terminal block connector and lift straight up away from the board.
5. Referring to the table below and the labels on the circuit board for terminal connections,
connect each wire to the terminal block.
6. To terminate a wire, loosen the screws in the terminal block and then insert the wire into the
terminal opening. Tighten the screw to secure the wire in place.
7. Repeat the procedure until all wires are in place.
8. After all terminations have been made, remove the excess cable from the enclosure.

P3 I/O Connections (Refer to Figure No. 6)

SERIAL INTERFACE
P3 Board Label
TXD
RXD
GND
+9-24 DC

P2 Power Connections (Refer to Figure No. 6)

P2
Power
10
Function
Transmit
Receive
Ground
9V DC to 24V DC
P2 Board Label
+9V
GND
CHGND
+BAT
-BAT
P2
Figure No. 6
8400-M022-O1 Rev K  Model 190 Indicator
INPUT/OUTPUTS
P3 Board Label
IN1
OUT1
OUT2
OUT3
COMMON
Function
9V DC
Ground
Chassis Ground
+ (Plus) Battery
- (Negative) Battery
Function
Input 1
Output 1
Output 2
Output 3
Common
P3
I/O
P3