Cardinal 190 Installation, Technical And Operation Manual page 38

Table of Contents

Advertisement

CAL3= – Third Calibration Weight
The display will show CAL3=. This is the second of two calibration weights. This
1.
weight could be ZERO (NO LOAD) or the TEST WEIGHTS (TEST LOAD).
Press the TARE  key to view the current setting.
2.
If the second calibration weight is to be ZERO (NO LOAD), press the TARE  key.
3.
If the first calibration weight is to be the TEST WEIGHTS (TEST LOAD), use the
4.
UNITS/ and Fn/ keys to input the value of the test weights.
Place the weights on the scale platform, then press TARE  key.
5.
Starting at the left and proceeding right, a series of dashes will appear on the display.
6.
Next, starting at the left and proceeding right, the dashes will disappear, after which the
7.
display will show: CAL4=.
CAL4= – Fourth Calibration Weight
The display will show CAL4=. This is the fourth of the calibration weights. This weight
1.
could be ZERO (NO LOAD) or the TEST WEIGHTS (TEST LOAD). If the fourth
calibration weight is not going to be used, press the UNITS/ key to skip CAL4= and
advance to SetgC? prompt. Otherwise, proceed to the next step.
Press the TARE  key to view the current setting.
2.
If the fourth calibration weight is to be ZERO (NO LOAD), press the TARE  key.
3.
If the fourth calibration weight is to be the TEST WEIGHTS (TEST LOAD), use the
4.
UNITS/ and Fn/ keys to input the value of the test weights.
Place the weights on the scale platform, then press TARE  key.
5.
Starting at the left and proceeding right, a series of dashes will appear on the display.
6.
Next, starting at the left and proceeding right, the dashes will disappear, after which the
7.
display will show: CAL5=.
CAL5= – Last Calibration Weight
The display will show CAL5=. This is the fifth of the calibration weights. This weight
1.
could be ZERO (NO LOAD) or the TEST WEIGHTS (TEST LOAD). If the fifth calibration
weight is not going to be used, press the UNITS/ key to skip CAL5= and advance to
SetgC? prompt. Otherwise, proceed to the next step.
Press the TARE  key to view the current setting.
2.
If the fifth calibration weight is to be ZERO (NO LOAD), press the TARE  key.
3.
If the fifth calibration weight is to be the TEST WEIGHTS (TEST LOAD), use the
4.
UNITS/ and Fn/ keys to input the value of the test weights.
Place the weights on the scale platform, then press the TARE  key.
5.
Starting at the left and proceeding right, a series of dashes will appear on the display.
6.
Next, starting at the left and proceeding right, the dashes will disappear, after which the
7.
display will show: SetgC?.
30
8400-M022-O1 Rev K  Model 190 Indicator