Accessing Setup
1. With the 190 turned ON, press the Fn/ and UNITS/ key simultaneously.
2. Hold both keys until the display changes to SetUP.
3. Release the keys to begin setup.
4. Press the UNITS/ key to step to the beginning point of each setup section.
SEtUP
A-d
CAL
SetgC
Sio
Print
F SPAn
Hi rES
LoCoUt
option
fUnC
CoLors
5. If you press the TARE key at the SetUP prompt, you may proceed through to the next
section (up to and including Colors) by pressing the TARE key.
6. To exit setup, press the Fn/ key with any of the menu selections displayed.
NOTE: Except for the SEtUP prompt, the prompts displayed for each section are
different if you push the UNITS/ key to step through the prompts instead of
pressing the TARE key to proceed through the section.
For example, if you press the UNITS/ key with the SEtUP displayed, the next
prompt displayed will be A-d. If you step through the setup prompts by pressing the
TARE key, the next prompt displayed will be A-d?.
Also, at a prompt with the ? displayed, you must press the TARE key again to
proceed with that section.
To skip the section and advance you to the next menu selection, press the TARE
key twice.
14
See Note Below
A-d?
CAL?
SetgC?
Sio?
Print?
FSPAn?
HirES?
LCoUt?
Opt?
fUnC?
CoLor?
8400-M022-O1 Rev K Model 190 Indicator
Setup Mode (starts at USA= prompt)
Analog to Digital Filtering (starts at dFLt= prompt)
Calibration (starts at CAL1= prompt)
Set Gravity Constant (starts at CALgC= prompt)
Serial Input/Output (starts at bAUd= prompt)
Print Tab Settings (starts at Port= prompt)
Fine Span Adjustment
Display High-Resolution Weight
Key Lockout Feature Setup
Configuration for Indicator Option Boards
Function Setup
Display Colors Setup