Analog to Digital Filtering

A-d
With A-d displayed, press the TARE  key. The display will change to dFLt=. Proceed to
the dFLt= parameter.
A-d?
With A-d? displayed, press the TARE  key. The display will change to no. Press the
Fn/ key to toggle to yes and press the TARE  key. The display will change to dFLt=.
Proceed to the dFLt= parameter.
Otherwise, to skip the Analog to Digital Filtering setup, press the TARE  key to advance to
the CAL?. prompt.
dFLt= (Digital Filtering)
With dFLt= displayed, press the TARE  key to show the current setting. If the setting
displayed is acceptable, press the TARE  key to save it.
Otherwise, use the Fn/ key to toggle to a new setting and press the TARE  key to save
it. Allowable settings are 0, 1, 2, or 3. Note, that if you select 3 (Custom Filtering) two
additional prompts will be displayed.
dFLt=
Filter Level = 2, Break Range = 1
0
Filter Level = 6, Break Range = 12, Sample Rate = 2
1
Filter Level = 20, Break Range = 12, Sample Rate = 1
2
3
CUSTOM FILTERING
NOTE: The prompts, F= (Filter Level) and b= (Break Range) will only be displayed if you
selected 3 (Custom Filtering) for the dFLt= (Digital Filtering) prompt.
Digital Filtering (dFLt=) selections 0, 1, and 2 have fixed factory settings for Filter Level,
Break Range, and Sample Rate.
F= (Filter Level)
The filter level is a number from 1 to 99 that corresponds to the level of filtering with 1 being
the least and 99 being the greatest. Press the TARE  key to show the current setting. To
accept the setting displayed, press the TARE  key again to save it.
Otherwise, use the Fn/ and UNITS/ keys to enter a new setting (1 to 99) and press the
TARE  key to save it.
8400-M022-O1 Rev K  Model 190 Indicator