Calibration "C" Numbers

IMPORTANT! If any components have been changed that affect calibration
and/or your scale is used in a commercial application and must be "Legal for
Trade" you cannot use the "C" numbers to re-calibrate.
The "C" numbers are displayed only during the Test mode operation by pressing the
Fn/ key then the UNITS/ key. The "C" numbers are shown at the end of the test operation
and each number is displayed for approximately 4 seconds, allowing you to record them.
Each number may be up to three (3) digits in length. By recording these numbers, you will be
able to return the indicator to its present calibration settings without using test weights simply
by entering the "C" numbers.
With CAL1= displayed, press UNITS/ key.
1.
At C1= prompt, press TARE  key to show the current value of C1 number.
2.
If the C1= value displayed is acceptable, press TARE  key again to save it.
3.
Otherwise, use Fn/ and UNITS/ keys to enter a new C1= value and press the
4.
TARE  key.
At C2= prompt, press TARE  key to show the current value of C2 number.
5.
If the C2= value displayed is acceptable, press TARE  key again to save it.
6.
Otherwise, use Fn/ and UNITS/ keys to enter a new C2= value and press the
7.
TARE  key.
At C3= prompt, press TARE  key to show the current value of the C3 number.
8.
If the C3= value displayed is acceptable, press TARE  key again to save it.
9.
Otherwise, use Fn/ and UNITS/ keys to enter a new C3= value and press
10.
TARE  key.
At C4= prompt, press the TARE  key to show the current value of the C4 number.
11.
If the C4= value displayed is acceptable, press the TARE  key again to save it.
12.
Otherwise, use Fn/ and UNITS/ keys to enter a new C4= value and press
13.
TARE  key.
The display will change to show: SetgC?.
14.
8400-M022-O1 Rev K  Model 190 Indicator