Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 2
C AU T I O N translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by A CAUTION notice denotes a United States and international hazard. It calls attention to an copyright laws.
About Flow Measurements USB Interface Warning and Error Indications Warning indication Error indication Low battery indication Power Indications Installing and Replacing the Cartridge Installing the cartridge Swapping or replacing the cartridge with ADM Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 4
Replacing the Batteries Upgrading the Firmware/Retrieving Certificate of Calibration Using the Flow Meter Stand A Note About Backpressure Technical Support Other Features Safety and Regulatory Certification Class A EMC certification Agilent CrossLab CS ADM Flow Meter Operation Manual...
ADM Flow Meter Introduction Introduction Figure 1. The Agilent CrossLab CS ADM flow meter front screen buttons. The Agilent ADM flow meter is controlled with two buttons on the front of the instrument directly below the screen, as shown Figure 1.
• To set a reference value in Split mode, briefly press and release Enter/Clear/Toggle button. • To change the settings in Info mode, press and hold the Enter/Clear/Toggle button for two seconds. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Figure 2. The initial screen display for the Agilent ADM flow meter. With the unit powered on, press and hold the Power/Mode button for three seconds to power down the unit. Figure 3. The shutdown screen for the Agilent ADM flow meter. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Use only a 5 V DC, 0.5 A USB power source and the WARNING supplied USB cable to run this device. • Do not use the ADM flow meter as safety equipment. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 9
USB de 0,5A, DC 5V y el cable de alimentación suministrado. • No utilice el flujómetro ADM como equipo de seguridad. • 仅使用 5v DC、0.5A USB 电源和随附的 USB 电缆为本设 警告 备充电。 • 请勿将 ADM 流量计用作安全设备。 Agilent CrossLab CS ADM Flow Meter Operation Manual...
Installing and Replacing the Cartridge on page 31. The flow meter is powered by three AA alkaline batteries. Power can also be supplied via the micro USB port located on the side of the unit. Agilent CrossLab CS ADM Flow Meter Operation Manual...
* When verified at ambient temperature of 20 to 24 °C. † Damage to the instrument may occur if it is exposed to temperatures greater than 50 °C, leading to inaccurate results. Agilent CrossLab CS ADM Flow Meter Operation Manual...
0.5 to 9.99 0.01 10.0 to 99.9 100 to 750 When the battery strength indicator shows that the battery strength is low, replace the batteries. See Replacing the Batteries on page 33. Agilent CrossLab CS ADM Flow Meter Operation Manual...
In this mode, the volumetric flow rate is corrected taking into account the current atmospheric pressure and ambient temperature. The flow rate is displayed in sccm (standard cubic centimeters per minute) referenced to one atmosphere of pressure and 0 °C. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Brightness screen is displayed. Press and hold the Enter/Clear/Toggle button for two seconds to toggle between High and Low options. The screen brightness changes based on the selected option. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 15
Auto off screen is displayed. Press and hold the Enter/Clear/Toggle button for two seconds to toggle between On and Off options. The Auto off function changes based on the selected option. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 16
04 No Cart Error No cartridge Install a valid cartridge. detected 05 No MF Info Warning No mainframe Contact Agilent information Technical Support 06 Auto Recover Warning Auto-recovery mode Reboot the flow meter Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 17
12 Ver Error Firmware version Upgrade firmware to Mismatch mismatch with latest version cartridge version 13 Invalid Cart Error Unable to identify Contact Agilent cartridge types Technical Support Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 18
When the cartridge is expired, the rectangular bar is fully filled and Expired! is blinking in the rectangular bar to alert user. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 19
Enter/Clear/Toggle button to cycle through available screens until the Mainframe screen is displayed. The Mainframe screen displays information about the mainframe. This includes mainframe model number (MDL), serial number (SN) and firmware version number (Ver). Agilent CrossLab CS ADM Flow Meter Operation Manual...
Wait for the reading to stabilize (1 to 2 seconds), then observe the reading. Low flows (less than 1 mL/min) may take up to 5 seconds for NOTE the reading to stabilize Agilent CrossLab CS ADM Flow Meter Operation Manual...
If the cartridge is unused, the values will be listed as nan (not a number). Press and hold the Enter/Clear/Toggle button to clear the previously stored values. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 22
Repeat step 2 to store additional values to the subsequent slots. Up to four values can be stored. To clear the stored values: • Hold down the Enter/Clear/Toggle button. All of the stored values are cleared from memory. Agilent CrossLab CS ADM Flow Meter Operation Manual...
1:1. Disconnect the flexible tubing from the GC column flow then connect it to the split vent or the gas flow that you want to compare. The display shows the ratio directly. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 24
Adjust for the desired ratio by adjusting the flow of the split vent, or the other reference flow, until the required ratio is displayed. To clear the stored reference value, press the Enter/Clear/Toggle button. Agilent CrossLab CS ADM Flow Meter Operation Manual...
As the measured gas or the flow meter itself is warmed, the amount of water vapor increases, and the readings can be much higher than the true flow rate. Agilent CrossLab CS ADM Flow Meter Operation Manual...
USB port. Data from the flow meter can be collected by connecting the USB port to your computer with the included USB cable. The flow meter USB driver will be needed for the instrument to communicate with your PC. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 27
When the flow meter is connected to PC software, the buttons on the flow meter are disabled. This is indicated by an icon to the left of USB icon on the screen. Agilent CrossLab CS ADM Flow Meter Operation Manual...
RTC battery is low, and so on. Error indication When a hardware fault occurs, an error icon is displayed on the screen. This example shows when a cartridge is not detected in the flow meter. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Low battery indication When the batteries need to be replaced, a low battery indication is displayed on the screen. When power is too low for operation, the flow meter will automatically shut down. Agilent CrossLab CS ADM Flow Meter Operation Manual...
The flow meter can be powered by batteries or through the USB port. When powered by batteries, a battery icon will appear on the screen. When powered through the USB port, a USB icon will appear on the screen. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Installing and Replacing the Cartridge Installing the cartridge The CrossLab CS Mainframe is a platform that can accept the ADM Flow Meter Cartridge P/N G6692A and Electronic Leak Detector Cartridge P/N G6694A. If you purchased the leak detector cartridge standalone,...
Page 32
ADM Flow Meter Swapping or replacing the cartridge with ADM flow meter cartridge Insert cartridge of ADM flow meter in to the CrossLab CS Mainframe body. Hand tighten the two captive thumbscrews. Install new flexible tubing. Tubing Agilent CrossLab CS ADM Flow Meter Operation Manual...
Press and hold the Power/Mode button for three seconds to power off the flow meter. Remove the battery cover by pressing in on the grooved thumb pad on the rear of the flow meter and sliding the battery cover off. Agilent CrossLab CS ADM Flow Meter Operation Manual...
Page 34
Reinstall the battery cover. Position the battery cover over the batteries and slide onto the flow meter until it locks in place. Briefly press the Power/Mode button to power on the flow meter. Agilent CrossLab CS ADM Flow Meter Operation Manual...
It is also possible to retrieve a digital copy of your cartridge's Certificate of Calibration in PDF form though the ADM flow meter's website. Please visit http://www.agilent.com/chem/ADMflowmeter for instructions. Agilent CrossLab CS ADM Flow Meter Operation Manual...
The flip down wire stand will give you a better viewing angle when the flow meter is placed above bench level. Pull the stand down until it is in the locked position. Stand Agilent CrossLab CS ADM Flow Meter Operation Manual...
Split ratio mode with continuous split flow rate reading • NIST traceable calibration • Compatible with dry gases • Volumetric and mass flow measurements • USB communication • Replaceable flow meter cartridge - individually computer calibrated Agilent CrossLab CS ADM Flow Meter Operation Manual...