Agilent Technologies CrossLab CS Operation Manual page 19

Electronic leak detector
Hide thumbs Also See for CrossLab CS:
Table of Contents

Advertisement

Leak Detector
Info mode
Table 3
Error and warning code descriptions
Code
01 No Cal Info
02 No Verif
03 No Setting
04 No Cart
05 No MF Info
06 I2C Fail
07 No Fltr Date
08 Filter Exp
09 Low RTC Bat
10 Zero Fail
11 Need Zero
12 Ver
Mismatch
13 Invalid Cart
Agilent CrossLab CS Electronic Leak Detector Operation Manual
Type
Description
Error
No calibration
information
Error
Leak detector is not
verified
Error
Leak detector
settings missing
Error
No cartridge
detected
Warning
No mainframe
information
Error
Hardware
communication
failure
Warning
No filter date
information
Warning
Filter expired
Warning
RTC (real time clock)
backup battery low
Error
Baseline correction
error
Error
Large voltage drift
because voltage out
of range
Error
Firmware version
mismatch with
cartridge version
Error
Unable to identify
cartridge types
Troubleshooting
Contact Agilent
Technical Support
Contact Agilent
Technical Support
Contact Agilent
Technical Support
Install a valid cartridge
Contact Agilent
Technical Support
Reboot the leak
detector
Contact Agilent
Technical Support
Renew the filter and
reset the filter new date
Contact Agilent
Technical Support
Reboot the leak
detector
Press the Clear button
in Detection mode
Upgrade the firmware
to latest version
Contact Agilent
Technical Support