Agilent Technologies CrossLab CS Operation Manual page 22

Electronic leak detector
Hide thumbs Also See for CrossLab CS:
Table of Contents

Advertisement

Leak Detector
Info mode
22
Agilent CrossLab CS Electronic Leak Detector Operation Manual
Filter date reset
The Reset Filter Date screen allows the filter date to be updated
when a new filter has been installed. The Reset Filter Date
screen is hidden and can only be prompted when the Yes option
in the Filter Exchanged screen is selected.
To change the Reset Filter Date option:
1
In the Reset Filter Date screen, press and hold the
Enter/Clear/Toggle button for two seconds to toggle
between Yes and No option.
2
With the desired option highlighted, briefly press and
release the Enter/Clear/Toggle button to select the
highlighted option.
After the Yes option is selected, the Filter New date is reset to
the current Pacific time, and the Filter Due date is set to one
year after the Filter New date.