Quick Setup Menu - Rice Lake SCT-2200 Advanced Series Technical Manual

Hide thumbs Also See for SCT-2200 Advanced Series:
Table of Contents

Advertisement

1.3

Quick Setup Menu

Press
to turn the instrument on. Enter the quick setup menu by pressing
C
Press
1
DIV.DEC
2
CEL.CAP
Press , and
Press , and
3
CEL.SEN
Press , and
4
DEAD.LD
5
RANGE
Press , and
Quick calibration of zero.
6
ZERO
Press ,
7
WEIGHT
8
ADC.MVV
Check the load cell signal in millivolts.
9
AN.OUT
10
INPUTS
11
OUTPUT
12
CLOCK
= Optional Analog Model
When settings are complete press
and
to set decimals and minimum division.
to set the total capacity of the load cells.
to set the mV/V sensitivity of the load cell.
to set the weight value present on the load cell.
to set the scale capacity.
and
to set the weight value used for the calibration.
MODE
Set the analog output on net or gross weight.
AO MIN
Set the minimum value of the analog output.
AO ZER
Set the analog output when the scale shows zero weight.
AO MAX
Set the maximum value of the analog output.
INP.01
Set the key to be emulated by the input.
INP.02
INP.01
INP.02
until the indicator displays SAVE?. Press
Pressing any other key exits the setup and discards changes.
Figure 1-3. Quick Setup Menu
as the firmware version displays.
FUNC
0 NONE
1 GROS
2 NET
3 PCS
4 GRO.0
5 NET.0
6 MOTI
23 K.PR
25.K.MOD
26 K.C
27.K.ZER
28.K.TAR
29 ERR
Set the output normal state.
NO/NC
ONSTAT
Set the output switching condition.
RL.IST
Enable or disable the hysteresis.
ENAB.TM
Enable time to disable the output.
DELAY
Enable time to enable the output.
to save set up and return to the weigh mode.
Disabled.
Set point on the gross weight.
Set point on the net weight.
Set point on the pieces.
Gross weight at zero.
Net weight at zero.
Active on instability.
Active when
is pressed.
Active when
is pressed.
Active when the C key is pressed.
Active when
is pressed.
Active when
is pressed.
Error indication.
Introduction
5