SCT-2200
Quick Setup Guide
To enter the setup menu, turn the instrument on and briefly press the mode key (
Press the print key (
) in the setup menu to view/select the parameter choices.
Press
1
DIV.DEC
Press
2
CEL.CAP
Press
3
CEL.SEN
Press
4
DEAD.LD
Press
5
RANGE
Quick calibration of zero.
6
ZERO
Press
7
WEIGHT
Check the load cell signal in millivolts.
8
ADC.MVV
9
AN.OUT
Optional Analog
Model Parameter
10
INPUTS
11
OUTPUT
Default Factory Calibration
The instrument has a default calibration with the following
features:
Capacity = 10,000 kg
Load cell sensitivity = 2.000mV/V
Divisions = 1
Theoretical Calibration
Use parameters 1,2,3,4 to perform a theoretical calibration.
Calibration Using Known Weight
Use parameters 1,5,6,7 to perform a standard calibration
with a known weight. For parameters 6 & 7, press
wait for ERNOT, STORE?. Then press
For parameter 7, apply test weight before pressing
Setpoint Programming
In the weigh mode, press and hold
setpoint programming.
When settings are complete press
November 7, 2018
and
to set decimals and minimum division.
,
and
to set the total capacity of the load cells.
,
and
to set the mV/V sensitivity of the load cell.
,
and
to set the weight value present on the load cell.
,
and
to set the scale capacity.
,
and
to set the weight value used for the calibration.
MODE
Set the analog output on net or gross weight.
AO MAX
Set the maximum value of the analog output.
AO ZER
Set the analog output when the scale shows zero weight.
AO MIN
Set the minimum value of the analog output.
INP.01
Set the key to be emulated by the input.
INP.02
OUT.01
OUT.02
and
to accept.
.
to access output's
until the indicator displays SAVE?. Press
Pressing any other key exits the setup and discards changes.
FUNC
0 NONE
1 GROS
4 GRO.0
5 NET.0
6 MOTI
23 K.PR
25.K.MOD
27.K.ZER
28.K.TAR
29 ERR
30.NET.T
Set the output normal state.
NO/NC
ONSTAT
Set the output switching condition.
RL.IST
Enable or disable the hysteresis.
ENAB.TM
Enable time to disable the output.
DELAY
Enable time to enable the output.
to save set up and return to the weigh mode.
) when the software version displays.
Key
Function
In setup: scroll forward through parameters
In numeric input: decreases the digit to be modified
ZERO
In setup: scroll back through the functions
In numeric input: increases the digit to be modified
TARE
In setup: quickly position the first step of a menu
In numeric input: selects the digit to be modified, from
MODE
left to right
In setup: enter into a parameter or to confirm a setting
In numeric input: confirms the entry made
PRINT
Turn the instrument on and off
In setup: exit a step without confirming the setting
ON/OFF
In numeric input: clears the present value
Disabled.
Setpoint for gross weight.
2 NET
Setpoint for net weight, even with no tare.
3 PCS
Setpoint for pieces.
Gross weight at zero.
Net weight at zero.
Active on instability.
Active when
Active when
is pressed.
26 K.C
Active when the
Active when
is pressed.
Active when
is pressed.
Error indication.
Setpoint for net weight when tare exists.
is pressed.
key is pressed.
PN 184974 Rev A