Quick Setup Guide
Key Functions
Key
Function
In setup: scroll forward through parameters
In numeric input: decreases digit to be modified
ZERO
In setup: scroll back through the functions
In numeric input: increases digit to be modified
TARE
Setup Menu
1. Press the ON/OFF key ( ) to turn the instrument on.
2. Press the mode key (
3. Press the print key (
Press
and
DIV.DEC
Press
,
and
CAPAC
Press
,
and
CEL.CAP
Press
,
and
CEL.SEN
(If more than one cell, set to average mV/V of the cells)
Press
,
and
DEAD.LD
Quick calibration of zero.
0.CALIB
For new acquisition of zero, with scale empty. Establishes new zero reference.
ZERo
Apply calibration weight. Press
SPAN
Check the load cell signal in millivolts.
ADC.uV
AN.oUT
Optional Analog
Model Parameter
INPUTS
OUTPUT
FLD.BUS
485
When settings are complete press
Note
weigh mode. Pressing any other key exits the setup and discards changes.
September 10, 2019
Key
In setup: quickly position the first step of a menu
In numeric input: selects digit to be modified, from
MODE
left to right
In setup: enter into a parameter or to confirm a
setting
PRINT
In numeric input: confirms the entry made
) when the firmware version displays to enter the setup menu.
) in the setup menu to view/select the parameter choices.
to set decimals and minimum division.
to set the scale capacity.
to set the total capacity of the load cells.
to set the mV/V sensitivity of the load cell.
to set the dead load present on the load cell.
,
and
to enter applied calibration weight.
MODE
Set the analog output on net or gross weight.
AO MAX
Set the maximum value of analog output.
AO ZER
Set the analog output when scale shows zero weight.
AO MIN
Set the minimum value of analog output.
INP.01
Set the key to be emulated by the input.
INP.02
OUT.01
OUT.02
no
hub
primar
yes
set.Add
Press
,
and
baud
Press
and
until the indicator displays SAVE?. Press
Function
Set the output normal state.
NO/NC
FUNC
no
yes
to set number of additional units.
to set baud rate
Key
Function
Turn the instrument on/off
In setup: exit a step without confirming the setting
ON/OFF
In numeric input: clears the present value
Default Factory Calibration
The instrument has a default calibration with the
following features:
Capacity = 10,000 kg
Load cell sensitivity = 2.000mV/V
Divisions = 1
Theoretical Calibration
Use parameters div.dec, capac, cel.cap, cel.sen,
dead.ld to perform a theoretical calibration.
Calibration Using Known Weight
Use parameters div.dec, capac, zero, span to
perform a standard calibration with a known weight.
For parameters zero & span, press
for ERNOT, STORE?. Then press
For parameter span, apply test weight before
pressing
.
Setpoint Programming
In the weigh mode, press and hold
put's setpoint programming.
GROS
Setpoint on the gross weight.
NONE
Disabled.
Setpoint on the net weight.
NET
SCA.Add
profib
ethip
num.sca
Profin
bus.typ
Eth.Cat
canopn
dev.net
mb.tcp
to save set up and return to
and wait
to accept.
to access out-
PN 184974 Rev B

Advertisement

loading
Need help?

Need help?

Do you have a question about the SCT-2200 and is the answer not in the manual?

Questions and answers

Summary of Contents for Rice Lake SCT-2200

  • Page 1 SCT-2200 Quick Setup Guide Key Functions Function Function Function In setup: scroll forward through parameters In setup: quickly position the first step of a menu Turn the instrument on/off In numeric input: decreases digit to be modified In numeric input: selects digit to be modified, from...
  • Page 2 Specifications subject to change without notice. Rice Lake Weighing Systems is an ISO 9001 registered company. 230 W. Coleman St. • Rice Lake, WI 54868 • USA U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319...